Articles

QUS of phalanx, DXA, BMD, Osteoporosis

Abstract

DXL of calcaneus is a portable method for BMD. This study determined a cut off point for DXL in osteoporosis diagnosis. In 510 healthy postmenopausal women, BMD of axial regions with DXA (DPX-MD, GE,Lunar Corp, Madison, WI) and heel with DXL (Demeteck- Sweden), measured. The agreement of two methods and cut off point for DXL in defining osteoporosis, obtained. DXA found osteoporosis in 34.3% and in DXL in 26.1% of cases. Agreement (Kappa) was 0.407 for spine and 0.347 for femur. T-score = -1.8 for spine and T-score = -2.2 for femur were the cut off points of DXL in diagnosis of osteoporosis (sensitivity=84% specificity=60%) and (sensitivity=84% and specificity=70%), respectively. Area under curve for regions were 0.807 (P=0.000) and 0.859 (P=0.000), respectively. These results mean DXL can not be used as a replacement for DXA, but it may can be used as a screening method for osteoporosis.
Files
IssueVol 33 No Supple 1 (2004) QRcode
SectionArticles
Keywords
DXL

Rights and permissions
Creative Commons License This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.
How to Cite
1.
A Soltani, R Hafezi, A Keshtkar, Z Hamidi, *B Larijani. QUS of phalanx, DXA, BMD, Osteoporosis. Iran J Public Health. 1;33(Supple 1):86-89.